Ethicon Endo Harmonic Hand Piece
- The Ethicon Endo Harmonic Hand Piece is designed to convert electrical energy from a compatible Harmonic Generator to mechanical motion for the instrument blades. This Hand Piece is intended for use with a compatible Harmonic Generator.
- The Hand Piece is permanently attached to a cord which connects to the front of the Generator
- These are compatible with Ethicon Endosurgery Gen11 Generator & are meant for use with all harmonic ultrasonic devices
- The Hand Piece is a re-useable instrument with a limited service life
- The Hand Piece is programmed with a counter to limit the service life to 95 procedures. The generator will give a Hand Piece error after 95 procedures are completed.
- The number of activations during a procedure is not limited, and the counter will not log a procedure until the Hand Piece is unplugged from the generator or the generator is powered down.
- Indications:
- The Harmonic Hand Piece, when used in conjunction with the Harmonic Instruments, is indicated for soft tissue incisions when bleeding control and minimal thermal injury are desired
- The instruments can be used as an adjunct to or substitute for electrosurgery, lasers, and steel scalpels
- Contraindications:
- The instruments are not indicated for incising bone
- The instruments are not intended for contraceptive tubal occlusion